Source voltage and measure current
Не провереноFour-quadrant voltage source with simultaneous current measurement. Supports DC and pulsed output modes.
Команды
SOUR:VOLTSOUR:VOLT:RangeSENS:CURR:RangeSOUR:VOLT:ILIMRanges
- voltage
- 0 – 40.4 V
- current
- 0 – 3.03 A
- power
- 0 – 40.4 W
Prerequisites
- · Verify instrument identity, firmware version, and calibration status.
- · Set current compliance before enabling voltage output.
- · Use 4-wire (remote sense) connection for accurate DUT voltage.
Границы безопасности
- Maximum floating voltage is 250 V DC. Never exceed between any terminal and earth ground.
- Set current compliance before enabling output. Compliance must be below DUT damage threshold.